TO-600在線式氧氣分析儀的詳細資料:
Model: | TO 600 |
Style: | Online |
Measuring Range: | 0.001 ppb- 100.0 % |
Units: | %/ ppm/ ppb, Auto Unit |
Digits: | 4 |
Accuracy: | > Highest Calibration Point, ±2% of Full Scale < Highest & > High Points, ±2% of Highest Point < High & > Middle Points, ±2% of High Point < Middle & > Low Points, ±2% of Middle Point < Low & > Lowest Points, ±2% of Low Point < Lowest Point, ±2% of Low Point |
Response Time: | T95 ≤ 5 sec. |
Analog Output: | 4/20mA or 0-10V (Optional) |
Digital Output: | RS232 & USB Disk, or RS485 (Optional) |
Alarm Relay: | None |
Calibration@ 1 l/min: | 5 Points (Standard) |
Warming Up: | > 200 sec. (Suggestion) |
Ambient temperature: | -5°C to +45°C |
Trend Chart: | Yes |
Analog Output: | 4/ 20 mA, High/ Low Point adjustable |
Customer Calibration: | Yes |
Software Upgradeable: | Yes |
Recording Interval: | ≥5S, Selectable |
Sampling Pump: | Yes |
Power: | 100-240VDC, 50/60Hz |
氧氣分析儀的使用流程
一、安裝與準備工作
選擇安裝位置:
安裝在氣體流動穩定、無劇烈震動和溫度波動處,靠近監測源以減少延遲。
避免長距離或彎曲管道連接,減少氣阻和延遲,確保樣氣代表性。
防爆型儀器(如ExdIICT6等級)需符合易燃易爆環境的安全要求。
連接管道與電源:
進氣管道需密封良好,防止漏氣;樣氣流量需符合規格(如0.2-1L/min)。
電源電壓通常為24VDC,信號線需連接至DCS/PLC系統或數據采集端。
預處理系統配置:
可選配預處理套件(如過濾器、干燥器),用于去除雜質、冷凝水等,避免傳感器污染。
二、啟動與校準
開機自檢:
通電后等待儀器完成自檢程序,檢查傳感器、顯示屏及通信狀態。
部分型號(如AGA3000)支持遙控器操作,無需開蓋即可標定或調零。
校準操作:
零點校準:通入零點氣體(如氮氣或無氧氣體),按說明書調整零點基準。
量程校準:使用標準氧氣濃度氣體(如20.9%O?或定制濃度)進行線性校準,確保測量精度。
電化學傳感器需定期校準(如每月一次),順磁式傳感器穩定性較高,校準周期可延長。
參數設置:
設置采樣時間間隔(如每秒一次)、報警閾值(上下限)及通信協議(如4~20mA/RS485)。
部分儀器支持自動/手動模式切換,工業現場通常采用自動循環監測。
三、實時監測與數據處理
動態監測:
儀器實時顯示氧氣濃度(如LCD或OLED屏幕),部分型號可顯示濃度曲線。
數據可通過信號輸出至控制系統,用于聯動報警或工藝調整。
報警處理:
當氧氣濃度超出預設范圍(如低于安全限或高于爆炸極限)時,儀器會觸發聲光報警,需立即排查原因(如設備故障、工藝異常)。
數據分析與記錄:
定期導出監測數據,生成報表(如日報表、趨勢圖),分析氧氣濃度變化規律。
記錄維護日志(如校準時間、傳感器更換周期)以便追溯。